AS Autosampler Operator’s Manual
164
Doc. 065051-03 1/08
5.12 Removing the Sample or Prep Syringe Valve
1. On the
MENU of SCREENS
,
press
5
and
3
to go to the
PLUMBING
CONFIGURATION
screen.
2. Move the cursor to the
SAMPLE SYRINGE CONFIG
field or the
PREP SYRINGE
CONFIG
field (depending on which syringe requires initialization).
3. Press
Select
Δ
to select the
INITIALIZE
option.
4. Press
Enter
.
The syringe plunger moves to the
initial position and the syringe motor
shuts off.
5. Unscrew and remove the syringe
drive connecting screw and the two
screws on the syringe valve (see
6. Pull the valve and syringe assembly
straight out to remove it.
7. Holding the syringe at the top fitting,
unscrew it from the valve.
8. Set the syringe aside in a safe place.
Figure 5-31. Removing the Syringe Valve
Syringe Drive
Connecting Screw
(Remove)
Valve Screws
(Remove)