AS Autosampler Operator’s Manual
154
Doc. 065051-03 1/08
Step 2:
After clicking
Test
, adjust the
X-position
and/or
Y-position
values until
the needle is centered in the inject port.
Step 3:
•
Select the syringe speed and diverter valve position (if installed) that will
provide the maximum backpressure delivered to the needle seal during
use. This backpressure must not exceed 690 kPa (100 psi). Refer to
for guidelines on selecting syringe speeds.
•
After selecting the syringe speed and the diverter valve position (if
installed), click
Test
. The AS dispenses 100
μ
L of fluid into the port.
•
Monitor the top of the inject port for fluid. If fluid appears, wipe up the
fluid immediately. Then, increase the needle’s
z-axis height
and click
Move
to move the needle down.
Note
: Each digit in the
z-axis height
value moves the needle 0.127 mm (0.005 in).
Step 4:
•
After clicking
Check
, the AS dispenses 500
μ
L of fluid into the port. This
larger amount of fluid provides a more thorough check than the 100
μ
L
that is dispensed by clicking
Test
in Step 3.
•
After clicking
Finish
, the new calibration values are displayed on the
panel. Write these values on the label inside the AS front door.
5. If you are performing this procedure after replacing the needle seal assembly,
go on to calibrate the inject port volume (see
Achten Sie beim Betrieb mit geöffneter Tür auf unerwartete
Bewegungen des Probenarms.
Increase the z-axis height value only until no more leaks occur. If the
z-axis height is too large, the needle will hit the bottom of the port
during operation and can damage the needle or port.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...