3 • Operation and Maintenance
Doc. 065051-03 1/08
65
•
If you enter an injection volume equal to or larger than the
installed loop size, the AS performs a full-loop (or large-loop)
injection.
•
If you enter an injection volume smaller than the loop size, the
AS performs a partial-loop or partial-loop, limited-sample
injection.
3. Enter the cut volume
in the Chromeleon or Chromeleon Xpress
program (see
SYSTEM PARAMETERS
screen
NOTE The cut volume is used only for partial-loop or
partial-loop, limited-sample injections. For full-loop
or large-loop injections, the cut volume is ignored.
See
for details about cut volumes.
Figure 3-17. Program Wizard: Specifying the Cut Volume