3 • Operation and Maintenance
Doc. 065051-03 1/08
71
Setting Up a Full-Loop Injection
1. On the
SYSTEM PARAMETERS
screen (see
), select
NORMAL
in the
SAMPLE MODE
field.
2. In the Chromeleon or Chromeleon Xpress sequence (see
or in the front panel schedule (see
), enter an injection
volume equal to or greater than the volume of the loop installed on
the injection valve.
Full-Loop Injection Notes
•
Entering an injection volume equal to or greater than the sample loop
size sets up the AS for a full-loop injection. The actual volume of
sample injected equals the volume of the sample loop installed.
•
If the entered injection volume is more than four times the loop size,
the entire injection volume is delivered during loading of the loop.
For example, if a 100
μ
L loop is installed and the injection volume is
600
μ
L, then 350
μ
L is flushed through the loop, 100
μ
L is loaded
into the loop, and 150
μ
L remains before the loop.