3 • Operation and Maintenance
Doc. 065051-03 1/08
89
5. Specify the injection volume and the cut volume required for the type
of injection to be performed. The cut volume specifies a volume of
sample to be “cut” (or omitted) from each side of the sample during
the injection process. To set up a full-loop injection, see
For a partial-loop injection, see
, and for a partial-loop
limited-sample injection, see
Example Program for Sequential Mode
The following example Chromeleon or Chromeleon Xpress program
shows the AS commands for sample loading and injecting when the
autosampler is in sequential mode.
Sampler.AcquireExclusiveAccess
;The timebase acquires
exclusive control of the AS, to
prevent another timebase from
attempting to control the AS
DiverterValve.Position_1
;Divert flow to system #1
0.000
Sampler_InjectValve.LoadPosition
;Switch the inject valve on
system #1 to Load
Load
;Load sample onto the loop
Wait
CycleTimeState
;Wait for the end of the
specified Cycle Time
Sampler_InjectValve.InjectPosition
;Switch the inject valve on
system #1 to Inject
Inject
;Sweep the sample onto the
column
Wait
InjectState
;Wait for the injection to finish
5.00
BeginOverlap
;Begin preparing the next
sample
Sampler.ReleaseExclusiveAccess
;Release control; another
timebase can now acquire
exclusive control of the AS
30.00
End
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...