5 • Service
Doc. 065051-03 1/08
133
5.3.1
Removing the Existing Syringe
1. Using a flathead screwdriver,
unscrew and remove the
syringe drive connecting
screw (see
2. Hold the syringe at the top
and bottom and unscrew it
from the valve.
3. Remove the white Teflon
washer from the syringe
valve port.
Note:
To remove
the washer, push a piece of
0.8-mm (0.031-in) OD PEEK
tubing into the center of the
washer and twist to dislodge
the washer.
4. Install a new white Teflon
washer (P/N 061315) into the
new syringe.
Figure 5-4. Removing the Syringe
Syringe Drive
Connecting
Screw (Remove)