AS Autosampler Operator’s Manual
88
Doc. 065051-03 1/08
Setting Up a Sequential Mode Injection
The AS must be equipped with the sequential injection option
(P/N 063294) in order to perform sequential injections. The following
setup steps are required:
1. Follow the instructions in
to install the sequential
injection diverter valve, connect the AS inject port tubing to the
diverter valve, and connect the diverter valve to the injection valves
on two systems.
2. Calibrate the inject port volume of each system. This is the volume
from the inject port through the diverter valve to the injection valve
(see
3. On the
SYSTEM PARAMETERS
screen, select
SEQUENTIAL
in the
SAMPLE MODE
4. On the
PLUMBING CONFIGURATION
screen, enter the sample loop size
installed on each valve (see
Figure 3-27. System Parameters Screen: Sequential Mode
Figure 3-28. Plumbing Configuration Screen: Sequential Mode
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
SEQUENTIAL
FLUSH SYRINGE SPEED:
4
uL
S2 47
Help Prompt
PLUMBING CONFIGURATION
LOOP SIZE V1:
SAMPLE SYRINGE VOLUME:
PREP SYRINGE VOLUME:
SAMPLE SYRINGE CONFIG:
5
25
250
READY
uL
uL
mL
PREP SYRINGE CONFIG:
READY
uL
LOOP SIZE V2:
100
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...