AS Autosampler Operator’s Manual
114
Doc. 065051-03 1/08
•
[409] Non-recoverable motion error occurred in X-axis. Clear any
physical obstructions and then press any key to reset system.
Possible Causes:
An error occurred during X-axis movement of
the sampling arm. The DSP was not able to recover from the
error.
Actions:
1. Turn off the power.
2. Check for an obstruction in the autosampler compartment.
3. Check for freedom of movement in all three axes of the sampling
arm.
4. Turn on the power again. If the problem persists, contact Dionex.
•
[414] Specified volume is larger than size of sample syringe.
During normal operation (not during diagnostics), the syringe can
perform up to four multiple strokes to aspirate or dispense a volume
up to four times larger than the syringe size.
Possible Causes:
•
During a schedule or flush/prime operation, the specified volume
is more than four times the size of the sample syringe.
•
On the
LIQUID CONTROL
diagnostics screen (see
either the specified volume exceeds the size of the sample syringe
or the cumulative volume (when performing multiple actions)
exceeds the size of the syringe.
Actions:
•
Reduce the volume specified or, if the sample preparation option
is installed, use the prep syringe (not the sample syringe) for
certain operations. See
“Sample Preparation Option Functions”
for details.
•
On the
LIQUID CONTROL
screen, select a different operation or
reduce the volume specified.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...