5 • Service
Doc. 065051-03 1/08
129
5.2
Resetting the Usage Counters
The Chromeleon Predictive Performance feature monitors the lifetime of
serviceable parts by counting the number of times a part has been used since it
was installed. After replacing or servicing a monitored part, reset the usage
counter, so that Chromeleon can then begin monitoring use of the new part. See
for more information about Predictive Performance.
Chromeleon monitors usage of the following AS parts: injection valve, diverter
valve, XYZ sampling needle arm, sampling needle assembly, needle seal
assembly, sample syringe, and prep syringe.
NOTE Reset the counter for the XYZ sampling needle arm
after lubricating or replacing the arm. Reset other usage
counters after replacing the part.
You can reset usage counters from either the AS Wellness panel in Chromeleon or
Chromeleon Xpress (see
) or on the front panel
USAGE LOG SCREEN