AS Autosampler Operator’s Manual
98
Doc. 065051-03 1/08
Example Backpressure Calculations for a Dual ICS-3000 System
The example tubing IDs and lengths for a dual ICS-3000 system shown in
will be used to calculate the total backpressure produced for each
system. See the following tables for details.
Prep Syringe Size
Prep Syringe Speed
Flow Rate (mL/min)
250, 500, and 1,000
μ
L
1
0.03
2
0.09
3
0.24
4
0.60
5
1.20
2.5, 5, and 10 mL
a
1
1.80
2
2.40
3
4.80
4
7.20
5
15.00
Table 3-12. Flow Rates Provided by the Prep Syringe
a. In general, if you are using a 5 or 10 mL syringe, select a syringe speed of 1 or 2.
Figure 3-35. Example Tubing IDs and Lengths for a Dual ICS-3000 System
1
2
P
P
P
Needle Seal Line
Diverter Valve
Diverter Line #1
W
L
P
C
L
S
Inject Valve #1
Inject Valve #2
W
L
P
C
L
S
Diverter Line #2
0.010-in ID, 46 cm
S
0.013-in ID, 38 cm
0.013-in ID, 64 cm
Inject Port
Needle
0.010-in ID
12 cm
25
µ
L Loop
5
µ
L Loop
0.020-in ID
15 cm