AS Autosampler Operator’s Manual
106
Doc. 065051-03 1/08
•
[237] Memory is full. Cannot save additional method or schedule.
The current method or schedule occupies all available memory.
Possible Causes:
Too many complex methods or schedules have
been saved.
Actions:
To free memory for future use, delete a method or
schedule that is no longer used.
•
[250] DSP does not acknowledge.
The DSP (digital signal processor) did not respond to a command
from the CPU.
Possible Causes:
Communication problem between the DSP and
the CPU.
Actions:
•
Turn the power off and then back on.
•
If the error message appears again, contact Dionex.
•
[331] Method does not exist.
The schedule contains a method that does not exist.
Possible Causes:
Incorrect method number entered into the
schedule.
Actions:
•
Enter the method number of an existing method in the schedule.
•
Create a new method and save it to the method number indicated
in the schedule.