E • TTL and Relay Control
Doc. 065051-03 1/08
321
E.2 Controlling TTL and Relay Outputs
The AS provides two TTL outputs and two relay contacts for control of functions
in external devices, such as an integrator. The relay outputs can be used to switch
any low-voltage control. Switched current must be less than 200 mA and 60 V
peak blocking. The relay-contact closures are normally open. When the relay is
closed, current flows to the connected device.
The TTL outputs are normally at 5 volts. Setting a TTL output to 0 volts turns on
the action in the connected device.
After connecting the TTL and Relay outputs (see
), turn on and off the
output states in Direct Control from the
DETAIL
), or in
Schedule Control during the timed events portion of the method. You can also use
Chromeleon or Chromeleon Xpress to control the outputs.
•
To turn on a TTL or relay output, set the corresponding output field on the
DETAIL
screen or method
TIMED EVENTS
screen to 1 (on).
•
To turn off a TTL or relay output, set the corresponding output field to 0 (off).
Figure E-4. Detail Status Screen - Direct Control of TTL and Relays
Relay loads in excess of 200 mA or with included power supplies over
60 V may damage the relay drivers on the CPU.
Help Prompt
IDLE
VIAL#:
TIME:
INJ#:
VOL:
METHOD:
TRAY:
TRAY:
LOOP:
20 °C
25
100
2 mL
uL
uL
min
TTL1
TTL2
RLY1
RLY2
0
1
0
0
LOCAL
DETAIL
DIRECT CONTROL
INJ VLV L
NORMAL FULL
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...