B • Installation
Doc. 065051-03 1/08
215
6. If the sample preparation option is installed, follow these steps to
prime the lines from each installed reagent reservoir (see
a. Select the reagent reservoir, enter
2000
μ
L
in the
Prime Volume
field, and press
Prime
.
b. Repeat
for each additional reagent reservoir.
7. When priming is complete, there are often small air bubbles in the top
area of the syringe. Remove the bubbles before proceeding. See
for instructions.
Suggested Priming Volumes
Sample Syringe Size
Suggested Priming Volume
≤
1,000
μ
L
2,000 to 2,500
μ
L
5,000
μ
L
9,999
μ
L
a
a. The maximum value that can be entered.
10,000
μ
L
19,998
μ
L
b
b. Run two priming cycles of 9,999
μ
L each.
Table B-2. Suggested Priming Volumes
(Sample Preparation Option Not Installed)
Reservoir
Prep Syringe Size
Suggested Priming Volume
Flush
250
μ
L
2,000
μ
L
500
μ
L
2,000
μ
L
1,000
μ
L
4,000
μ
L
2.5 mL
5,000
μ
L
5 mL
9,999
μ
L
a
a. The maximum value that can be entered.
10 mL
19,998
μ
L
b
b. Run two priming cycles of 9,999
μ
L each.
A, B, C, and D
c
c. Repeat the priming procedure for each reagent reservoir.
All
2,000
μ
L
Table B-3. Suggested Priming Volumes
(Sample Preparation Option Installed)
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...