D • Front Panel Control
Doc. 065051-03 1/08
309
19. Press
8
to go to the
DIAGNOSTICS MENU
. Press
5
to go to the
LIQUID
CONTROL
screen.
20. Select
VIAL#: INJ
and press
Enter
.
The needle moves into the inject port.
21. Position the cursor in the
from
field and select
FLUSH
.
22. Position the cursor in the
SYRINGE
field and select
SAMPLE.
23. Position the cursor in the
ACTION
field and select
FILL
.
24. Press
Enter
.
The syringe fills.
25. Select
SYRINGE SPEED: 5
. This speed ensures a fast enough flow rate to
detect a leak.
26. Select
ACTION:
EMPTY
and press
Enter
.
27. Repeat the
FILL
and
EMPTY
actions until the new needle seal tubing is filled.
28. Once the tubing is filled and fluid reaches the needle, empty the syringe again
and monitor the top of the inject port for fluid.
If fluid appears, wipe up the leak immediately. Then, lower the needle’s Z
position (see the next step).
29. To lower the needle’s Z position:
a. Press
Menu
,
Menu
,
5,
and
6
to return to the
INJECTION PORT ALIGNMENT
screen.
b. Increase the
Z ADDRESS
field value two steps and press
Enter
.
c. Repeat
through
, above, and check again for leaks.
Figure D-17. Liquid Control Screen
Help Prompt
LIQUID CONTROL
from
VIAL#:
ASPIRATE:
SYRINGE:
ACTION:
DISPENSE:
1
100
100
SAMPLE
ASPIRATE
uL
uL
NEEDLE
LOAD
46
NEEDLE HEIGHT:
SYRINGE SPEED: 4
INJ VALVE:
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...