AS Autosampler Operator’s Manual
136
Doc. 065051-03 1/08
2. Move the cursor to the
SAMPLE
SYRINGE CONFIG
field or the
PREP
SYRINGE CONFIG
field (depending on
which syringe is to be removed).
3. Press
Select
Δ
to select the
INITIALIZE
option.
4. Press
Enter
.
5. The syringe plunger moves down a
small amount. This is its initial
position (see
Figure 5-7. Initializing the Syringe
5.3.5
Resetting the Syringe Usage Counter
After replacing the syringe, reset the syringe usage counter from either
the AS Wellness panel in Chromeleon or Chromeleon Xpress or on the
front panel
USAGE LOG SCREEN
5.4
Replacing the Sampling Needle Assembly
Replace the sampling needle assembly after about 6,200 cycles of use (the cycle
count increments each time the needle is inserted into the inject port). When the
AS is operating under Chromeleon or Chromeleon Xpress control, a reminder
message is displayed in the Audit Trail when the number of cycles reaches this
usage limit. When the AS is operating under front panel control, the
USAGE LOG
Initial
Position
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...