C • Display Screens
Doc. 065051-03 1/08
241
Notes on Reagent Volume Ranges
•
If the source is a reservoir, the allowed range is 100 to 50,000
μ
L. The
default is 2,000
μ
L if the syringe volume is less than 5,000
μ
L, and
10,000
μ
L if the syringe volume is 5,000
μ
L or 10,000
μ
L.
•
If the source is a vial, the range depends on the volume of the sample
syringe:
NOTE The
REAGENT FLUSH
and
REAGENT PRIME
operations
are performed
after
the autosampler loads the sample
onto the concentrator column. See
for
details about concentrator injections.
Sample Syringe
Volume (
µL)
Reagent Flush
Range (
µL)
Default (
µL)
100 to 1,000
100 to 1,000
1,000 (any size vial)
5,000
100 to 4,700
1,000 (1.5 mL vial)
4,000 (10 mL vial)
10,000
100 to 8,000
1,000 (1.5 mL vial)
8,000 (10 mL vial)