2 • Description
Doc. 065051-03 1/08
21
•
The AS automatically detects the type of tray installed.
2.3.6
Injection Valve Options
The AS can optionally have one or two injection valves installed inside
the autosampler compartment. The injection valves are six-port,
electrically-activated Rheodyne valves. The valves have two operating
positions: Load and Inject (see
Eluent flows through either the Load or Inject path, depending on the valve
position:
•
In the Load position, sample flows from the AS inject port line, through the
valve, and into the sample loop, where it is held until injection. Eluent flows
from the pump, through the valve, and to the column, bypassing the sample
loop. Excess sample flows out to waste.
•
In the Inject position, sample is swept to the column for analysis. Eluent flows
from the pump, through the sample loop, and on to the column, carrying the
contents of the sample loop with it.
Figure 2-8. Injection Valve Flow Schematics
(1)
L
P
(2)
C
(3)
L
(4)
(5)
S
(6)
W
To Waste
Sample In
To Column
From Pump
Sample
Loop
LOAD POSITION
To Waste
Sample In
From Pump
Sample
Loop
INJECT POSITION
To Column
Concentrator
Column
(optional—
replaces loop)
(1)
L
P
(2)
C
(3)
L
(4)
(5)
S
(6)
W
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...