AS Autosampler Operator’s Manual
224
Doc. 065051-03 1/08
B.5.2
Creating Sequences, Programs, and Quantification
Methods
Follow these guidelines when creating sequences, programs (PGM file),
and quantification methods (QNT file) for simultaneous injection:
•
Create one sequence that includes calibration standards for both
systems and the unknown samples.
•
Create a PGM file for the unknowns that includes commands for
controlling both systems.
•
Calibrate each system separately by creating separate calibration
PGM files.
•
Create a single QNT file that includes the components from both
systems.
Creating a Method File for Systems Sharing an AS
For simultaneous injection analyses, create a single method file with all
components for each detector channel included in one component list.
Use the
Duplicate Column
command on the Peak Table, Amount Table,
and Peak Tracking pages of the method to create two extra
Ret. Time
columns; associate each column with a different channel.
NOTE To add a column, right-click in the Ret. Time
column and select the Columns>Duplicate Column
command on the context menu.
For detailed instructions, refer to Chromeleon Help.
In the following example Peak Table, the
Peak Name
column includes
the components from two detector channels: one for cations and one for
anions. The
Ret. Time ECD_1
column is associated with the cation
detector channel and displays only the cations. The
Ret. Time ECD_2
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...