4 • Troubleshooting
Doc. 065051-03 1/08
113
•
[390] Received invalid command from DSP.
Possible Causes:
Communication between the CPU and the DSP
was corrupted.
Actions:
Turn off the power and then turn on the power again. If
the message reappears, contact Dionex.
•
[391] Schedule error. Specified schedule does not exist.
Possible Causes:
The schedule selected to run has not been
saved.
Actions:
Create the new schedule or select an existing schedule.
•
[398] Battery backed up RAM failed data validation test. Setting
system parameters to defaults and initializing schedule and
method database.
Possible Causes:
•
Temporary short on CPU board
•
Dead battery on CPU board
•
Malfunctioning power supply
Actions:
If a temporary short is suspected, the AS can still be
operated. You will need to re-enter any method and schedule
information and customized system setup parameters. If the
problem persists, contact Dionex.
•
[408] Recoverable motion error occurred in X-axis
Possible Causes:
An error occurred during X-axis movement of
the sampling arm. The DSP was able to recover from the error
and continue the motion.
Actions:
Occasional occurrences of this error can be ignored. If
the problem persists, contact Dionex.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...