Chapter 2 Before Use
2-38
2.15.2 Precautions When Using Bias-T
When connecting an external device such as a Bias-T to the output
connectors of the BERTWave, if the output signal includes any DC
voltage, variations in the output of the DC power supply or load may
change the level of the output signal, risking damage to the internal
circuits.
CAUTION
Do not connect or disconnect any external devices
while DC voltage is impressed.
Only switch DC power sources ON and OFF when all
equipment connections have been completed.
Figure 2.15.2-1 Bias-T Connection Example
DUT
DC power source
50
Bias-T
Coaxial cable
Coaxial cable
Set output ON/OFF after
completing connections.
To protect DUT and PPG
Do not connect/disconnect while
DC voltage impressed.
BERTWave
Be sure to connect the ground structure (like frame ground) and the ground
terminal with a ground wire.