9.2 Performance Test for Error Detector
9-11
9
Performance T
est
15. Adjust the PPG1 Amplitude so that the Eye/Pulse Scope
amplitude is 50±1 mV.
16. Connect the PPG1 Data Out and ED1 Data In connectors using
a coaxial cable. (Refer to (b) in Figure 9.2.2-1.)
17. Touch [Start] at All Measurements.
18. Record the error count ER for the ED Result after measurement
is complete.
19. Connect coaxial terminators to the PPG1 Data Out and ED1
Data In connectors. (Refer to (a) in Figure 9.2.2-2.)
20.
Attach a 20 dB fixed attenuator to the PPG1 Data Out
connector.
21. Connect the PPG1 Data
Out and Ch A In connectors using a
coaxial cable.
22. Repeat steps 5 to 7.
23. Connect the PPG1 Data Out and ED1 Data In connectors
using a coaxial cable. (Refer to (b) in Figure 9.2.2-2.)
24. On the [ED] tab, set Data Input Condition to [Single-Ended
XData].
25. Repeat steps 10 to 12.
26. Connect the PPG1 Data Out and Ch A In connectors using a
coaxial cable. (Refer to (a) in Figure 9.2.2-2.)
27. Repeat steps 14 and 15.
28. Connect the PPG1 Data Out and ED1 Data In connectors
using a coaxial cable. (Refer to (b) in Figure 9.2.2-2.)
29. Repeat steps 17 and 18.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...