Chapter 3 Connecting with DUT
3-10
When the DUT is an optical transceiver (using clock recovery)
The trigger clock can be generated from the received signal when the
Option 023 is installed and the Option 054 is added.
1. Connect the CRU Out to Trigger Clk In using a coaxial cable.
2. Connect the PPG output to the DUT input connector using a coaxial
cable.
3. Connect the DUT output and O/E Data In using an optical fiber.
CAUTION
Check that the DUT optical output level does not exceed
the rated optical input level for O/E Data In.
There is a risk of damaging the O/E module if the optical
power exceeding the rated optical input level is applied to
the O/E Data In connector.