Appendix A Specifications
A-16
Table A.1.3-6 Vertical System (Cont’d)
Item
Specifications
Electrical input
(Cont’d)
Amplitude accuracy
*
3
±amplitude accuracy ±2% for reading value*
4
The following figure shows the amplitude accuracy after calibration.
0
5
10
15
20
25
0
50
100
150
200
250
300
350
400
Differences at Offset Values (mV)
A
m
p
lit
ud
e A
ccur
a
cy
(
±m
V
)
Scale=1 mV/div
Scale=15 mV/div
Scale=50 mV/div
Scale=100 mV/div
Scale=200 mV/div
Scale=250 mV/div
*3: After internal calibration
*4: Calculation example: At 400 mV amplitude reading value and 50 mV
offset voltage
Offset voltage
50 mV
Scale
50 mV/div.
1 level
200 mV
0 Level
–
200 mV
Differences at offset
values 250 mV
Differences at offset
values
150 mV
Amplitude
measurement
value
400 mVp-p
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...