Chapter 2 Before Use
2-22
Forced shutdown
Press the Power switch for 4 seconds or more.
The green Power switch lamp goes off, the sand by lamp lights orange,
and the power is turned off.
Notes:
Only use forced shutdown as an emergency operation when key,
mouse, and keyboard operations fail. A fault may have occurred
if the power cannot be turned off even by pressing the Power
switch for 4 or more seconds. Unplug the power cord from the
power outlet and contact your Anritsu Service and Sales Office
or agent.
If the power plug is removed while the panel access lamp is lit,
the data may not be saved correctly. Depending on the type of
data that was not saved correctly, sometimes the equipment
may not start correctly at the next power-on. Remove the power
plug after cutting the power.
The power may not turn on correctly if the power is turned off
and then turned on again immediately after the Stand by lamp
illuminates orange.
Wait at least 5 seconds after turning off the power before
turning on again.