7.1 EYE/Pulse Scope Screen
7-9
7
Measurin
g W
aveform
Item Description
Channel Math
[Off]: Displays waveforms for Channel A
and B separately.
[On]:
Calculates waveforms for Channel A
and B, and then that calculated
result is displayed as Channel A.
Scale Offset
Sets the level scale for Channel A and B.
Channel A/B
Tracking
[Off]: Sets scales for Channel A and B
separately.
[On]:
Sets scales for Channel A and B as
the same value.
Scale Sets
vertical
scale.
Offset Sets
vertical
offset.
Attenuation
Sets attenuation amount for external
attenuation.
Channel Offset
Sets Channel A and Channel B waveform
calculation method and level scale for
calculation results waveform.
Define Function
Sets calculation method between channels.
Scale
Sets vertical scale for calculation result
between channels.
Offset
Sets vertical offset for calculation result
between channels.