9.3 Performance Test for Sampling Oscilloscope
9-21
9
Performance T
est
9.3 Performance Test for Sampling Oscilloscope
This section describes the performance test item for the sampling
oscilloscope.
Amplitude accuracy
9.3.1 Required equipment
The following table shows the equipment required for the performance
test.
Before starting performance tests, warm up the BERTWave and
measuring instruments for at least 1 hour.
Table 9.3.1-1 Equipment for Performance Test
Product
Name
Required Performance
Recommended
Equipment
Pulse Pattern
Generator
Clock frequency:
10 GHz
Amplitude:
0.5
Vp-p
MP1800A,
MU181000A
MU181020A
(Anritsu)
DC Power
Source
Power voltage:
±2.5 V
Power current:
±50 mA
Setting accuracy:
1% or less
With current limiting function
2400
(Keithley
Instruments)