7.5 Calibrating
Sampling
Oscilloscope
7-23
7
Measurin
g W
aveform
After the following operations, CAL alarm is displayed on the screen.
When the CAL alarm is displayed, calibrate the level.
●
Change [Sampling Mode] in the Setup dialog box.
●
When Sampling Mode is [Pulse] or [Coherent Eye], change [Pattern
Length] in the Time dialog box.
CAL Alarm
Figure 7.5.1-3 CAL Alarm Display
7.5.2 Performing Self Test
Perform the self test in the following cases:
●
When the error message is displayed after level calibration referring to
●
When the noise voltage is over the specified value shown in Appendix A
when the signal is not input
●
When the waveform is not displayed, the displayed waveform
amplitude and bit cycle is different from the expected value, and the
BERTWave operation and measurement results are abnormal.
The application self test runs following items.
●
Power
●
Frequency
●
Amplitude
If the test result is normal, “passed” is displayed, on the other hand, if the
test result is abnormal, “failed” is displayed.
The self test procedures are as follows.
1. Check that no signals are input to the front-panel Ch A In, Ch B In,
Trigger Clk In connectors.
2. Touch [Setup] to open the Setup dialog box.
3. Touch the [Utilities] tab.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...