Chapter 9 Performance Test
9-4
9.1.3 Waveform
(1) Specification
Amplitude:
0.1 to 0.8 Vp-p
Accuracy:
(±20% of setting) ±20 mV
Data Crossing:
50±10%
(Amplitude: 0.8 Vp-p, Bit rate: 11.32 Gbit/s (12.5
Gbit/s when with the Option 092))
Rise/fall time:
28 ps
(Amplitude: 0.8 Vp-p, Percentage: 20-80%, Bit
rate: 11.32 Gbit/s (12.5 Gbit/s when with the
Option 092))
Jitter (RMS):
2 ps (Amplitude 0.4 Vp-p, 10.3125 Gbit/s)
(2) Setup
Sampling Oscilloscope
Sync Out
Trigger
Input
PPG1 Data Out
Input
Coaxial Terminator
Figure 9.1.3-1 PPG1 Data Out Waveform Test Connection Diagram