Chapter 5 Generating Pulse Signal
5-14
When monitoring the waveforms of two DUTs simultaneously, make the
following connections.
Figure 5.4-4 Setup for Simultaneous Monitoring of Two Channels
Set the Ch2 Reference Clock to Ch1 even when the measurement system
includes long-distance transmission.
Figure 5.4-5 Measurement System Including Long-distance
Transmission
Procedures:
1. Touch the screen size change button, and set to [Expand].
2. Touch the Reference CLK button.
The selection screen is displayed. The number of the displayed
buttons vary with the option configuration.
3.
Touch the reference clock button, referring to Table 5.4-1.
DUT
Sync
Out
Trigger Input
Option 012 or 014
DUT
Ch A
Input
Set Ch2 Reference
Clk to Ch1
Ch B
Input
Sampling
Oscilloscope
Sync
Out
Option 012 or
014
Sampling
Oscilloscope
Ch A
Input
Set Ch2 Reference
Clk to Ch1
Ch B
Input
Optical Transceiver
(Transmitter)
Optical Transceiver
(Receiver)
Long Fiber