Chapter 5 Generating Pulse Signal
5-26
5.8 Setting Sync Output
Set the signal type output from the front-panel Sync Out connector.
Sync Output connector outputs either a pulse synchronized to the data
generated by the PPG, or a pulse synchronized to the data received by the
ED, and outputs it at the front-panel Sync Out connector.
When measuring the bit error rate and waveform using another error
detector or sampling oscilloscope, the Sync Out signal is connected to the
measurement equipment.
The Sync Out Connector is a DC coupling.
The output of the Sync Output connector is input either to a sampling an
oscilloscope trigger to monitor the PPG waveform.
Data Connector Voltage
Time
Time
0
0
Sync Out Connector Voltage
2,4,8,16, or 64
Figure 5.8-1 Correspondence between Sync Output Setting and Output Voltage
Waveform at Connector using PPG or ED Rate Division
Data Connector Voltage
Time
Time
0
0
Sync Out Connector Voltage
Figure 5.8-2 Correspondence between Sync Output Setting and Output Voltage
Waveform at Connector using PPG 1/1
Data Connector Voltage
Time
Time
0
0
Sync Out Connector Voltage
Pattern Length
Figure 5.8-3 Voltage Waveform Output to Connector Corresponding to Sync Output
Setting for PPG Pattern Sync
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...