Chapter 3 Connecting with DUT
3-8
When the DUT is an optical transceiver:
When the Option 023 is selected, the output waveform of the optical
transceiver can be measured using the O/E converter.
Input the output of the PPG in the BERTWave to the DUT, and then
connect the DUT optical output to the O/E Data In converter.
1. Connect the DUT input terminal to PPG1
Data
Out/Data Out
using a coaxial cable.
If only one DUT input connect is conveniently available, connect to
the PPG1 Data Out. Also, connect the coaxial terminator included as
a standard accessory to the
Data
Out.
2. Connect the DUT optical output connector to O/E Data In using an
optical fiber.
3. Connect the Sync Out to Trigger Clk Input using a coaxial cable.
CAUTION
Check that the DUT optical output level does not exceed
the rated optical input level for O/E Data In.
There is a risk of damaging the O/E module if the optical
power exceeding the rated optical input level is applied to
the O/E Data In connector.