Chapter 7 Measuring Waveform
7-24
4. Touch [Application Test]. The self test dialog box is displayed while
the self test is running.
5. After the self test is completed, the result is displayed.
When the power supply test result is failed:
1. When the power voltage is displayed, record the value.
2. If the power supply test result shows “failed”, contact an Anritsu
Service and Sales office.
When the frequency test result is failed:
1. Check that no signals are input to the Trigger Clk In connector.
2. Perform the self test again.
3. If the frequency test result shows “failed”, calibrate the sampling
oscilloscope again. And then, perform the self test again. Nonetheless,
in cases where the test result fails, contact an Anritsu Service and
Sales office.
4. Perform the self test again.
5. Nonetheless, in cases where the test result fails, contact an Anritsu
Service and Sales office.
When the amplitude test result is failed:
1. Check that no signals are input to the Ch A In, Ch B In connectors.
2. Perform the self test again.
3. If the amplitude test result shows “failed”, calibrate the sampling
oscilloscope again.
4. Perform the self test again.
5. Nonetheless, in cases where the test result fails, contact an Anritsu
Service and Sales office.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...