5.6 Setting
Patterns
5-21
5
Generating Pulse Signal
PRBS
Pattern
Generation
Polynomial
Pattern
Length
Maximum
number of
contiguous 1s
Maximum
number of
contiguous 0
2^7–1 1+X
6
+X
7
127
7 6
2^9–1 1+X
5
+X
9
511
9 8
2^15–1 1+X
14
+X
15
32767
15 14
2^23–1 1+X
18
+X
23
8388607
23 22
2^31–1 1+X
28
+X
31
214748647
31 30
The ratio of 1s to the PRBS pattern length is 50%.
Programmable Pattern
The Programmable Pattern is the pattern defined by the pattern file.
The Programmable Pattern is used when the following patterns are
required:
Frame format pattern such as Ethernet and SDH
Pattern captured from communication circuit data
Specified test pattern
The restrictions when using the Programmable Pattern are:
The occurrence ratio of “1” is 50% with ±1bit tolerance.
The pattern length is 1305600 bits or less. When the file data length is
over 1305600 bits, the first 1305600 bits are used.
When using pattern files created for existing Anritsu products, use
Pattern A only when the pattern file type is Alternate.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...