Chapter 5 Generating Pulse Signal
5-16
5.5 Setting Output Waveform
The PPG output terminal is the PPG
Data
and Data connectors on the
front panel.
The connector block diagram is as shown below.
Amplitude
Control
Data
Data
Figure 5.5-1 Output Terminal Block Diagram
The connector is an AC coupling.
When it is necessary to impose DC voltage on an output waveform,
connect a bias-T.
CAUTION
The impedance of the Data Out and Data Out
connectors is 50
Ω
. Measurement may not be
performed correctly if a coaxial cable with another
impedance is used or if the impedance of the DUT is not
50
Ω
.
The amplitude of signal output to the Data Out and Data
Out connectors is 0.1 to 0.8 Vp-p. Check that the output
voltage amplitude does not exceed the DUT maximum
input specifications. If it does, connect an attenuator to
the output connector.