Chapter 6 Measuring Bit Error
6-22
Measurement method
The ED Result display can be refreshed either in real time (approx. 1 s
intervals) or when the progress bar reaches 100%. The display method is
set at Current of Gating.
[On]: Updates
measurement
results in real time.
[Off]: Updates measurement results when progress reaches 100% for
either Single or Repeat Gating Cycle.
Updates measurement results when measurement stops for
Untimed Gating Cycle.
The setting procedure of the measurement conditions is as follows:
1. Touch the Gating Cycle button and select the measurement method
as follows:
[Repeat]
[Single]
[Untimed]
2. When Gating Cycle is Repeat or Single, touch the text box under the
Gating Cycle and enter the numeric value.
The measurement cycle can be set from 1 second to 9 days 23 hours
59 minutes 59 seconds.
3. Touch [Current] to set the timing for the measurement result.
[On]: Updates
measurement results per 100 ms.
[Off]:
Displays measurement results per measurement results
or when measurement stops
4. Touch the measurement start/stop button, and set the display to
[Start].
The right indicator of the button is changed to green.
Measure is displayed at the status display.
The measurement processing rate is displayed.
When pattern synchronization is obtained, the SYNC Loss display is
erased.
When Gating Cycle is set to [Untimed], the measurement cycle progress is
displayed every 5 seconds.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...