7.1 EYE/Pulse Scope Screen
7-5
7
Measurin
g W
aveform
Tab Item
Description
General
Sampling Mode
Switches to Eye/Pulse/ Coherent Eye mode.
Fast Sampling Mode
Increases the eye pattern data sampling speed if set to [On].
This parameter can be set when Sampling Mode is Eye.
Number of Samples
Sets the number of samples.
Accumulation Type
Sets the accumulation method of the sampling data.
Limit Type
Selects the sampling end conditions from time, sampling count,
or frequency.
Time
Sets the sampling time.
Sample
Sets the number of samples.
Waveforms
Sets the waveforms.
Averaging
When measuring the pulse mode, set the averaging process
count.
CRU Input
Displays the clock recovery input connector when the Option
053, 054, or 055 is added. When the Option 055 is added, the
details selected for ED1 Data Input Condition are displayed.
Clock Recovery
Sets the frequency band for clock recovery when the Option 053,
054, or 055 is added.
The input connector is set at the Data Input Condition of the ED.
CRU Loop BW
Sets the bandwidth of the loop filter when Clock Recovery is set
to [>8.5G].
CRU Lock
When the Option 053, 054, or 055 is added, touch [Continue
Scan] to detect the clock recovery frequency.
Utilities
EYE/Pulse Shot
Touching [Capture] saves the screen file in the result window.
Inverse background
color
Sets the color of the save screen at EYE/Pulse Shot.
Waveforms Only
On: Saves only the waveform displayed in the Result window to a
file. Off: Saves the whole Result window to a file.
NEW Label
Adds a new label.
Delete Label
Deletes the displayed label.
Set Reference
Saves the displayed waveform as the reference trace.
Clear Reference
Deletes the reference trace.
Ref.Trace Channel
Sets the channel saved in the reference trace.
Temperature
Displays current temperature of the sampling oscilloscope and
the temperature when the level is calibrated.
Calibration
Calibrates the amplitude level of the sampling oscilloscope.
Application Test
Runs the self test for the sampling oscilloscope.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...