9.3 Performance Test for Sampling Oscilloscope
9-23
9
Performance T
est
(3) Procedure
1. Connect the Trigger Input and the clock output connector of the
PPG using a coaxial cable.
2. Set the clock output of the PPG as follows:
Amplitude:
0.5
Vp-p
Frequency:
10
GHz
3. Touch [Setup], and set as follows:
Sampling Mode:
EYE
Accumulation Type: None
Clock Recovery:
Off
Divide Ratio:
1
4. Touch [Time], and set as follows:
Unit:
UI
Bits On Screen:
2
Offset: 0
Data Clock Rate – Tracking: Off
Clock Rate:
10 000 000 kHz
Divide Ratio:
1
5. Touch [Amplitude], and set as follows:
Channel Math:
Off
Channel A/B Tracking:
Off
Scale – A:
50 mV/div
Offset – A:
0 mV
Attenuation – A:
0 dB
6. Touch [Measure], and set as follows:
Active Channel Selection:
Channel A
Measure Item:
Histogram
Histogram – Axis:
Amplitude
X1:
0
X2:
2
Y1:
–250
mV
Y2:
250
mV
7. Set the Current Limit of DC Power Source to 20 mA.
8. Set the DC voltage to 0 V.
9. Connect the DC Power Source to Ch A In.
Connect the center conductor of the coaxial cable to plus and the
shield to minus.
10. Touch [CH A] to set to On.
11. Touch [Sampling] to set to Run.
12. Record the average value (Mean) of the histogram measurement
result.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...