6.4 Setting Measurement Conditions
6-23
6
Measuring Bit Error
When SYNC Loss is displayed in red
Pattern synchronization has not been obtained; check the following:
The Test Pattern generated by the DUT matches the Test Pattern for
the error detector.
The Logic POS and NEG settings are correct.
A suitable threshold voltage has been set for the signal input to the
Data In connector and the Data In connector.
Figure 6.4-2 Examples of Improper Thresholds
Threshold
Threshold is not between level
0 and 1
Excessive noise close to
threshold level
Threshold
Threshold