
D.2 Mask file Specifications
D-5
App
endi
x
A
ppendix D
The following items are described in the mask file.
Table D.2-2 Details of Mask File
Item
Parameter to be set
Description
Version Masks
Version#1
Make sure that the version
is written.
Mask name
Mask name
Mask name displayed on the
measurement screen.
Mask type
Relative Mask
Fixed Mask
Relative: Sets coordinate of
amplitude direction by rate.
The Mask coordinate varies
with the amplitude of input
signal.
Fixed: Sets coordinate of
amplitude direction at the
fixed value. The Mask
coordinate does not vary
with the amplitude of input
signal.
Coordinates in
the top
TOP X Y X_min Y_min
X_max Y_max
Specifies each X and Y
coordinate in the area of
Top/Middle/Bottom.
As for X_min and Y_min, set
the coordinate at 100% of
Mask Margin.
As for X_max, and Y_max,
set the coordinate at –100%
of Mask Margin.
The coordinate can be set up
to 10 points. If it is not
used, set “–1”.
Coordinates in
the middle
MIDDLE X Y X_min
Y_min X_max Y_max
Coordinates in
the bottom
BOTTOM X Y Y_min
X_max Y_max
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...