Math Waveforms
4
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
87
The following figure illustrates aliasing. This is the spectrum of a 990 Hz
square wave, which has many harmonics. The sample rate is set to
100 kSa/s, and the oscilloscope displays the spectrum. The displayed
waveform shows the components of the input signal above the Nyquist
frequency to be mirrored (aliased) on the display and reflected off the
right edge.
Because the frequency span goes from
≈
0 to the Nyquist frequency, the
best way to prevent aliasing is to make sure that the frequency span is
greater than the frequencies of significant energy present in the input
signal.
FFT Spectral Leakage
The FFT operation assumes that the time record repeats. Unless there is
an integral number of cycles of the sampled waveform in the record, a
discontinuity is created at the end of the record. This is referred to as
leakage. In order to minimize spectral leakage, windows that approach
zero smoothly at the beginning and end of the signal are employed as
filters to the FFT. The FFT Menu provides four windows: Hanning, Flat
Figure 9
Aliasing
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...