296
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
19
Utility Settings
4
Connect a BNC cable (40 inches maximum) from the TRIG OUT
connector on the rear panel to the BNC barrel connector.
5
Press the
[Utility]
key; then, press the
Service
softkey.
6
Begin the Self Cal by pressing the
Start User Cal
softkey.
To perform hardware self test
Pressing
[Utility] > Service > Hardware Self Test
performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
•
After experiencing abnormal operation.
•
For additional information to better describe an oscilloscope failure.
•
To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the
oscilloscope's functionality. Hardware Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
To perform front panel self test
Pressing
[Utility] > Service > Front Panel Self Test
lets you test the front panel
keys and knobs as well as the oscilloscope display.
Follow the on- screen instructions.
To display oscilloscope information
Press
[Help] > About Oscilloscope
to display information about your
oscilloscope:
•
Model number.
•
Serial number.
•
Bandwidth.
•
Module installed.
•
Software version.
•
Installed licenses. See also
"Loading Licenses and Displaying License
Information"
on page 323.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...