I2C/SPI Triggering and Serial Decode
24
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
363
3
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the I
2
C signals
are being decoded.
4
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
Start Condition
— The oscilloscope triggers when SDA data transitions
from high to low while the SCL clock is high. For triggering purposes
(including frame triggers), a restart is treated as a start condition.
•
Stop Condition
— The oscilloscope triggers when data (SDA) transitions
from low to high while the clock (SCL) is high.
•
Missing Acknowledge
— The oscilloscope triggers when SDA data is high
during any Ack SCL clock bit.
•
Address with no Ack
— The oscilloscope triggers when the acknowledge
of the selected address field is false. The R/W bit is ignored.
•
Restart
— The oscilloscope triggers when another start condition occurs
before a stop condition.
•
EEPROM Data Read
— The trigger looks for EEPROM control byte value
1010xxx on the SDA line, followed by a Read bit and an Ack bit. It
then looks for the data value and qualifier set by the
Data
softkey
and the
Data is
softkey. When this event occurs, the oscilloscope will
trigger on the clock edge for the Ack bit after the data byte. This
data byte does not need to occur directly after the control byte.
SDA
SCL
Stop
Condition
R/
Start
Condition
Ack
Ack
Data
Address
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...