Measurements
14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
235
Each analog channel and the math waveform can be assigned unique
threshold values.
3
Press the
Type
softkey to set the measurement threshold to
%
(percentage of Top and Base value) or to
Absolute
(absolute value).
•
Percentage thresholds can be set from 5% to 95%.
•
The units for absolute threshold for each channel is set in the
channel probe menu.
•
When the
Source
is set to
Math: f(t)
, the threshold
Type
can only be set
to
Percent
.
4
Press the
Lower
softkey; then, turn the Entry knob to set the lower
measurement threshold value.
Increasing the lower value beyond the set middle value will
automatically increase the middle value to be more than the lower
value. The default lower threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from
5% to 93%.
5
Press the
Middle
softkey; then, turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper
thresholds. The default middle threshold is 50% or 1.20 V.
•
If threshold
Type
is set to
%
, the middle threshold value can be set
from 6% to 94%.
T I P
Absolute threshold hints
• Absolute thresholds are dependent on channel scaling, probe attenuation, and probe units.
Always set these values first before setting absolute thresholds.
• The minimum and maximum threshold values are limited to on-screen values.
• If any of the absolute threshold values are above or below the minimum or maximum waveform
values, the measurement may not be valid.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...