CAN/LIN Triggering and Serial Decode
22
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
341
•
Aliased bus values (undersampled or indeterminate) are drawn in pink.
•
Unknown bus values (undefined or error conditions) are drawn in red
with a "?" label.
•
Flagged error frames are drawn in red with an "ERR" label.
CAN Totalizer
The CAN totalizer provides a direct measure of bus quality and efficiency.
The CAN totalizer measures total CAN frames, flagged error frames,
overload frames, and bus utilization.
The totalizer is always running (counting frames and calculating
percentages) and is displayed whenever CAN decode is displayed. The
totalizer counts even when the oscilloscope is stopped (not acquiring
data). Pressing the
[Run/Stop]
key does not affect the totalizer. When an
overflow condition occurs, the counter displays
OVERFLOW
. The counters
can be reset to zero by pressing the
Reset CAN Counters
softkey.
Types of Frames
•
Active error frames are CAN frames in which a CAN node recognizes
an error condition during a data or remote frame and issues an active
error flag.
•
A partial frame occurs when the oscilloscope detects any error
condition during a frame that is not followed by an active error flag.
Partial frames are not counted.
Counters
•
The FRAMES counter gives the total number of completed remote, data,
overload, and active error frames.
•
The OVLD counter gives the total number of completed overload frames
and their percentage of the total number of frames.
•
The ERR counter gives the total number of completed active error
frames and their percentage of the total number of frames.
•
The UTIL (bus load) indicator measures the percentage of time the bus
is active. The calculation is done on 330 ms periods, approximately
every 400 ms.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...