Vertical Controls
3
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
69
For certain active probes, such as InfiniiMax probes, the oscilloscope can
accurately calibrate its analog channels for the probe. When you connect a
probe that can be calibrated, the
Calibrate Probe
softkey in the Channel
Probe Menu becomes active.
To calibrate one of these probes:
1
First, plug your probe into one of the oscilloscope channels.
This could be, for example, an InfiniiMax probe amplifier/probe head
with attenuators attached.
2
Connect the probe to the left side, Demo 2, Probe Comp terminal, and
the probe ground to the ground terminal.
3
Press the Channel on/off key to turn the channel on (if the channel is
off).
4
In the Channel Menu, press the
Probe
softkey.
5
In the Channel Probe Menu, the second softkey from the left is for
specifying your probe head (and attenuation). Repeatedly press this
softkey until the probe head selection matches the attenuator you are
using.
The choices are:
•
10:1 single- ended browser (no attenuator).
•
10:1 differential browser (no attenuator).
•
10:1 (+6 dB Atten) single- ended browser.
•
10:1 (+6 dB Atten) differential browser.
•
10:1 (+12 dB Atten) single- ended browser.
•
10:1 (+12 dB Atten) differential browser.
•
10:1 (+20 dB Atten) single- ended browser.
•
10:1 (+20 dB Atten) differential browser.
6
Press the
Calibrate Probe
softkey and follow the instructions on the
display.
N O T E
When calibrating a differential probe, connect the positive lead to the Probe Comp terminal
and the negative lead to the ground terminal. You may need to connect an alligator clip to
the ground lug to allow a differential probe to span between the Probe Comp test point and
ground. A good ground connection ensures the most accurate probe calibration.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...