Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
2
Press the
Word Size
softkey; then, turn the Entry knob to select the
number of bits in a word.
3
Press the
Bit Order
softkey; then turn the Entry knob to select the bit
order, most significant bit first (MSB) or least significant bit first (LSB),
used when displaying data in the serial decode waveform and in the
Lister.
4
If the decode line does not appear on the display, press the
[Serial]
key
to turn it on.
5
If the oscilloscope is stopped, press the
[Run/Stop]
key to acquire and
decode data.
You can use the horizontal
Zoom
window for easier navigation of the
acquired data.
See Also
•
"Interpreting SPI Decode"
on page 377
•
"Interpreting SPI Lister Data"
on page 378
•
"Searching for SPI Data in the Lister"
on page 378
N O T E
If the setup does not produce a stable trigger, the SPI signal may be slow enough that the
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
softkey to set the trigger mode from Auto to Normal.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...