412
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
27
UART/RS232 Triggering and Serial Decode
•
When
#Bits
is set to 9 in the UART/RS232 Bus Configuration Menu,
the 9th (alert) bit is displayed directly to the left of the ASCII value
(which is derived from the lower 8 bits).
4
Optional: Press the
Framing
softkey and select a value. In the decode
display, the chosen value will be displayed in light blue. However, if a
parity error occurs the data will be displayed in red.
5
If the decode line does not appear on the display, press the
[Serial]
key
to turn it on.
6
If the oscilloscope is stopped, press the
[Run/Stop]
key to acquire and
decode data.
You can use the horizontal
Zoom
window for easier navigation of the
acquired data.
See Also
•
"Interpreting UART/RS232 Decode"
on page 413
•
"UART/RS232 Totalizer"
on page 414
•
"Interpreting UART/RS232 Lister Data"
on page 415
•
"Searching for UART/RS232 Data in the Lister"
on page 415
N O T E
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow
enough that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then
press the Mode softkey to set the trigger mode from Auto to Normal.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...