Digital Channels
6
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
111
3
Press the
D15 - D8
or
D7 - D0
softkey, then select a logic family preset or
select
User
to define your own threshold.
The threshold you set applies to all channels within the selected D15 - D8
or D7 - D0 group. Each of the two channel groups can be set to a
different threshold if desired.
Values greater than the set threshold are high (1) and values less than the
set threshold are low (0).
If the
Thresholds
softkey is set to
User
, press the
User
softkey for the
channel group, then turn the Entry knob to set the logic threshold. There
is one
User
softkey for each group of channels.
To reposition a digital channel
1
Make sure the multiplexed scale and position knobs to the right of the
key are selected for digital channels.
If the arrow to the left of the
[Digital]
key is not illuminated, press the
key.
2
Use the multiplexed Select knob to select the channel.
The selected waveform is highlighted in red.
3
Use the multiplexed Position knob to move the selected channel
waveform.
If a channel waveform is repositioned over another channel waveform,
the indicator at the left edge of the trace will change from
D
nn
designation (where nn is a one or two digit channel number) to
D
*. The
"*" indicates that two channels are overlaid.
Logic family
Threshold Voltage
TTL
+1.4 V
CMOS
+2.5 V
ECL
–1.3 V
User
Variable from –8 V to +8 V
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...