Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
16
Waveform Generator
Creating New Arbitrary Waveforms
The New Waveform Menu is opened by pressing
Create New
in the Edit
Waveform Menu.
To create a new arbitrary waveform:
1
In the New Waveform Menu, press
Initial Pts
; then, use the Entry knob to
select the initial number of points in the new waveform.
The new waveform will be a square wave with the number of points
you specify. The points are evenly spaced over the time period.
2
Use the
Frequency/Frequency Fine/Period/Period Fine
softkey to set the time
period bounding parameter (repetition frequency) of the arbitrary
waveform.
3
Use the
Amplitude
/
High-Level
and
Offset
/
Low-Level
softkeys to set the
voltage bounding parameter of the arbitrary waveform.
4
When you are ready to create the new arbitrary waveform, press
Apply
& Edit
.
Interpolate
Specifies how lines are drawn between arbitrary waveform points.
When enabled, lines are drawn between points in the waveform editor. Voltage levels
change linearly between one point and the next.
When disabled, all line segments in the waveform editor are horizontal. The voltage level of
one point remains until the next point.
Source
Selects the analog channel or reference waveform to be captured and stored to the arbitrary
waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
on page 258.
Store Source to Arb
Captures the selected waveform source and copy it to the arbitrary waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
on page 258.
Softkey
Description
N O T E
You can use the [Save/Recall] key and menu to save arbitrary waveforms to one of four
internal storage locations or to a USB storage device, and you can recall them later. See
"To
save arbitrary waveforms"
on page 272 and
"To recall arbitrary waveforms"
on page 275.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...