![Agilent Technologies InfiniiVision 3000 3024A MSO-X User Manual Download Page 144](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679144.webp)
144
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
Edge Trigger
The Edge trigger type identifies a trigger by looking for a specified edge
(slope) and voltage level on a waveform. You can define the trigger source
and slope in this menu. The trigger type, source, and level are displayed in
the upper- right corner of the display.
1
On the front panel, in the Trigger section, press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey, and use the Entry knob
to select
Edge
.
3
Select the trigger source:
•
Analog channel,
1
to the number of channels
•
Digital channel (on mixed- signal oscilloscopes),
D0
to the number of
digital channels minus one.
•
External
— triggers on the rear panel EXT TRIG IN signal.
•
Line
— triggers at the 50% level of the rising or falling edge of the AC
power source signal.
•
WaveGen
— triggers at the 50% level of the rising edge of the
waveform generator output signal. (Not available when the DC, Noise,
or Cardiac waveforms are selected.)
You can choose a channel that is turned off (not displayed) as the
source for the edge trigger.
The selected trigger source is indicated in the upper- right corner of the
display next to the slope symbol:
•
1
through
4
= analog channels.
•
D0
through
Dn
= digital channels.
•
E
= External trigger input.
•
L
= Line trigger.
•
W
= Waveform generator.
4
Press the
Slope
softkey and select rising edge, falling edge, alternating
edges, or either edge (depending on the selected source). The selected
slope is displayed in the upper- right corner of the display.
Summary of Contents for InfiniiVision 3000 3024A MSO-X
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Page 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Page 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Page 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Page 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Page 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...